• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LAG/COMP SCREW KIT 95/90; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LAG/COMP SCREW KIT 95/90; NAIL, FIXATION, BONE Back to Search Results
Model Number 71677095
Device Problems Fitting Problem (2183); Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case (b)(4).
 
Event Description
It was reported that, during an internal fixation surgery, the lag/comp screw kit 95/90 threads were not allowing the driver to engage the screw.Surgery was resumed, with a back-up device.It is unknown if there was any delay as consequence of this allegation.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device did not confirm the stated failure mode.A functional evaluation of the returned device could not confirm the stated failure mode.The device functioned as intended.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable causes could include but are not limited to user error or rough handling.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG/COMP SCREW KIT 95/90
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12823834
MDR Text Key280879202
Report Number1020279-2021-08195
Device Sequence Number1
Product Code JDS
UDI-Device Identifier03596010563316
UDI-Public03596010563316
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040212
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71677095
Device Catalogue Number71677095
Device Lot Number21GM18914
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-