OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problems Communication or Transmission Problem (2896); Device Sensing Problem (2917)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, a b30 scope communication error due to defective s-socket and damaged tank socket were noted.The probable cause of the malfunction was due to wear and tear.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported the evis exera iii xenon light source does not recognize any endoscope.In addition, an e30 error was displayed.There was no patient involvement, no harm or user injury reported due to the event.
 
Manufacturer Narrative
This follow up report is being submitted to include the device history record(dhr) review and results from the legal manufacturer investigation.The following sections were updated: b5, h2, h4, h6 and h10.The legal manufacturer performed the dhr review for this device and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.The root cause was unable to be identified however, it is surmised that the b30 error occurred because the scope socket was out of order.Olympus will continue to monitor complaints for this device.
 
Event Description
It was further reported that the problem was first identified during a gastroscopy procedure.The device was changed however, the intended procedure was not completed.There was no injury or harm caused to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12828223
MDR Text Key280901305
Report Number8010047-2021-14683
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-