Model Number N/A
Device Problems Incorrect Measurement (1383); Patient Device Interaction Problem (4001)
Patient Problem Laceration(s) (1946)
Event Type  Injury  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Initial reporter telephone number: (b)(6).Review of the most recent repair record determined the control bar was found to not be flush with the master blade and damage was noted on the head.; however, the repair was not completed because the customer requested a replacement device.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the device at first took too thin skin, then it took too thick skin.Occurred during surgery and there was unspecified harm and delay.The description of the event indicates additional grafts were needed.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key12830694
MDR Text Key280903253
Report Number0001526350-2021-01273
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)190314(10)64324727
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64324727
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received06/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-