• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CSF VENTRICULAR RESERVOIR, SIDE INLET, 18 MM; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CSF VENTRICULAR RESERVOIR, SIDE INLET, 18 MM; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR Back to Search Results
Model Number 44102
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem Inflammation (1932)
Event Date 05/17/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported the patient was implanted with the reservoir on (b)(6) 2021; however, on (b)(6) 2021, the patient was hospitalized due to cerebrospinal fluid leakage.On (b)(6) 2021, a revision surgery was performed due to inflammation in the area of the reservoir placement and suspicion of implant disfunction.The healthcare professional specialist examined the implant and discovered two holes of injection, and the damage on surface of the reservoir 1 mm in size.The healthcare professional conclusion was that it was not possible to define the reason of this damage appeared, whether it was cut with surgery lancet or it was a manufacturing defect.
 
Manufacturer Narrative
H2) additional information was received: lot number, udi, explanted date (d) initial reporter (e) manufacture date (h4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF VENTRICULAR RESERVOIR, SIDE INLET, 18 MM
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTR
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
5290 california ave
irvine CA 92617
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key12832780
MDR Text Key280929576
Report Number2021898-2021-00196
Device Sequence Number1
Product Code LKG
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
K833822
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number44102
Device Catalogue Number44102
Device Lot Number0220114769
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/18/2021
Date Device Manufactured05/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age16 YR
Patient SexFemale
-
-