MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL X-SMART PLUS; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL X-SMART PLUS; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number A103200000000
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2021
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.After inspection/investigation of device it was found to have various mechanical problems.Device required replacement of head assembly, cartridge, head cap and bearing retainer.Additional information regarding the patient outcome has been requested and will be submitted as it becomes available.
 
Event Description
It was reported that a x-smart plus would not hold files.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
After inspection/investigation of device it was found to have various mechanical problems.Device required replacement of head assembly, cartridge, head cap and bearing retainer.
 
Event Description
Additional information received indicating no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART PLUS
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key12843472
MDR Text Key282098690
Report Number8031010-2021-00352
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeKS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 01/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA103200000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/07/2021
Date Manufacturer Received10/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage Unknown
Patient Sequence Number1
-
-