Catalog Number A103200000000 |
Device Problems
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.After inspection/investigation of device it was found to have various mechanical problems.Device required replacement of head assembly, cartridge, head cap and bearing retainer.Additional information regarding the patient outcome has been requested and will be submitted as it becomes available.
|
|
Event Description
|
It was reported that a x-smart plus would not hold files.The event outcome is unknown as of this mdr evaluation.
|
|
Manufacturer Narrative
|
After inspection/investigation of device it was found to have various mechanical problems.Device required replacement of head assembly, cartridge, head cap and bearing retainer.
|
|
Event Description
|
Additional information received indicating no injury resulted.
|
|
Search Alerts/Recalls
|