• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE SERFAS ENERGY 90-S CRUISE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279401200
Device Problems Loss of or Failure to Bond (1068); Material Erosion (1214)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2021
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event Description
It was reported that the glue adhesive was flaking during the procedure.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: adhesive peels off probable root cause: non-conforming component.Poor assembly process.Misuse.Use error.The reported failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that the glue adhesive was flaking during the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERFAS ENERGY 90-S CRUISE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key12844945
MDR Text Key282103851
Report Number0002936485-2021-00648
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057034
UDI-Public07613327057034
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0279401200
Device Catalogue Number0279401200
Device Lot Number21034AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-