• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SECHRIST INDUSTRIES, INC SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3600; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SECHRIST INDUSTRIES, INC SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3600; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number 3600
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the returned device did not confirm the reported issue.The device functioned as expected and was within acceptance criteria.A dhr review found no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file no.(b)(4).
 
Event Description
Customer reported the gas supply alarm is not working.When only the oxygen line is attached (no air), and the output is closed (with our jet ventilator), the alarm does not sound.The issue was discovered at incoming inspection and no patient incident was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3600
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer (Section G)
SECHRIST INDUSTRIES, INC
4225 e. la palma ave.
anaheim CA 92807
Manufacturer Contact
majid mashayekh
4225 e. la palma ave.
anaheim, CA 92807
7145798309
MDR Report Key12845507
MDR Text Key282266279
Report Number2020676-2021-00034
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K992503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3600
Device Catalogue Number3600
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-