SECHRIST INDUSTRIES, INC SECHRIST AIR/OXYGEN GAS BLENDER MODEL 3600; MIXER, BREATHING GASES, ANESTHESIA INHALATION
|
Back to Search Results |
|
Model Number 3600 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Evaluation of the returned device did not confirm the reported issue.The device functioned as expected and was within acceptance criteria.A dhr review found no evidence that a manufacturing non-conformity contributed to the reported complaint, and the instructions for use were reviewed and determined to provide adequate instructions and warnings for the safe and effective use of the device.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file no.(b)(4).
|
|
Event Description
|
Customer reported the gas supply alarm is not working.When only the oxygen line is attached (no air), and the output is closed (with our jet ventilator), the alarm does not sound.The issue was discovered at incoming inspection and no patient incident was reported.
|
|