• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750469
Device Problems Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/22/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A biomedical engineer reported that a phacoemulsification handpiece heated after connecting to the console.The surgeon noted an unpleasant smell of burning.No system messages were displayed.The surgery was completed by using alternate handpiece.
 
Manufacturer Narrative
The phacoemulsification handpiece was not received for evaluation.A phaco handpiece manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The root cause of the reported event cannot be determined conclusively.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12849762
MDR Text Key281086227
Report Number2028159-2021-01373
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657504695
UDI-Public00380657504695
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K112425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750469
Device Lot Number13TM05
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-