Catalog Number 399.680 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported during a procedure on (b)(6) 2021, it was noticed the concave gouge had a fracture on the edge of the device.This report is for a gouge.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D4.Lot.D9.Date device returned to manufacturer.G1.Manufacturing site name and address.H4.Device manufacture date.H6 - codes updated to imdrf codes.Visual inspection: the gouge w/10 l205/90 was returned for analysis.Visual inspection of the returned device found a broken condition, the concave gouge presents a fracture on the edge of this piece.Dimensional inspection: dimensional inspection cannot be performed due to post manufacturing damage.Investigation conclusion: the reported condition of the complaint device (the gouge w/10 l205/90) is confirmed.Device is broken.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/ specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot - part: 399.680 lot: 5016533 manufacturing site: selzach supplier: (b)(4) release to warehouse date: 02 oct 2007 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|