RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Increase in Pressure (1491)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2021 |
Event Type
malfunction
|
Event Description
|
The customer called into technical support (ts) reporting that the device's event log displayed error code (1009) pressure regulation high message.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.The hospital respiratory therapist (rt) manually shut down the device and swapped out with a different device to use on the patient.No medical intervention provided to the patient.The customer evaluated the device with the assistance of the remote service engineer (rse).It was relayed that the rt department found pressure regulation high (ec 1009) error code logged.The biomed engineer evaluated the device and could not replicate or determine what the cause was, could not confirm reported problem.It was verified the proximal and machine pressure lines are not crossed.Discussed testing flows and pressures.The biomed customer could not replicate the reported issue.The device passed required performance verification tests per philip¿s standards and was put back into service.
|
|