In efforts to oxygenate the infant, inhaled nitric oxide (ino) was administered via ino delivery device, with high flow oxygen blender.The blender read 100%; however, the analyzer read 21%.Abgs revealed po2 of <30.There were multiple efforts to troubleshoot the setup and identify the issue, the ino vendor was called while respiratory care and neonatology attempted to increase oxygenation.The infant sustained a pneumothorax.The infant's health was declining rapidly and passed with parents.Vendor assessment of equipment was requested.It was determined the ino device was working properly but the o2 blender was not.It was found that the nut retaining the knob of the o2 blender was loose and not providing the friction fit required for adjustment of o2 concentration.
|