Model Number V3.0
Device Problems Use of Incorrect Control/Treatment Settings (1126); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that during cataract surgery the vitrectomy cutter of an ophthalmic operating console did not work.The surgery was completed by using another console.There was no patient impact.
 
Manufacturer Narrative
The company service representative, examined the system.And was unable to confirm or replicate the reported event.The system was then tested and met all product specifications.A system manufacturing device history record (dhr) review was performed, prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.The system was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12874586
MDR Text Key281273991
Report Number2028159-2021-01397
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657521609
UDI-Public00380657521609
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV3.0
Device Catalogue Number8065752160
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2013
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-