MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 BROACH CORAIL AMT 11; HIP INSTRUMENTS : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 BROACH CORAIL AMT 11; HIP INSTRUMENTS : BROACHES Back to Search Results
Model Number L20411
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/05/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
When broaching, the rasp became jammed in the patients femur which resulted in us splitting the proximal femur and cabling it to extract the said rasp.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will bereviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROACH CORAIL AMT 11
Type of Device
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12882066
MDR Text Key281319534
Report Number1818910-2021-26321
Device Sequence Number1
Product Code HTQ
UDI-Device Identifier10603295258131
UDI-Public10603295258131
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberL20411
Device Catalogue NumberL20411
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-