• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C Back to Search Results
Model Number EG-3870UTK
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2021
Event Type  malfunction  
Manufacturer Narrative
We checked the returned unit and confirmed that the ccd module defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the deflector wire stretched, the us probe failure, the us probe cut, the segment loosened, the light guide fiber bundle (lcb) cut and the us connector cable cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Video image failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
LINEAR ULTRASOUND VIDEO GASTROSCOPE 3.8C
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishisima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key12889206
MDR Text Key283620507
Report Number9610877-2021-50578
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K182004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG-3870UTK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2021
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-