DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Disconnection (1171)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to use the cs300 intra- aortic balloon pump (iabp) unit displayed iab disconnected error.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: the getinge field service engineer (fse) evaluated the cs300 intra-aortic balloon pump (iabp) unit and observed the unit displaying iab disconnected error.To fix the issue, the fse replaced the k6a vent valve, 24 vdc, 2-way no.Which caused the error and code 66 in the logs.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: the getinge field service engineer (fse) evaluated the cs300 intra-aortic balloon pump (iabp) unit and observed the unit displaying iab disconnected error.To fix the issue, the fse replaced the k6a vent valve, 24 vdc, 2-way no.Which caused the error and code 66 in the logs.The fse performed functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|