Catalog Number 312400 |
Device Problems
Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/05/2020 |
Event Type
malfunction
|
Event Description
|
Empty material (no solution).
|
|
Manufacturer Narrative
|
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.
|
|
Manufacturer Narrative
|
Based on no photograph or sample was received for analysis from the customer, the failure mode could not be confirmed.The ampoule is made of glass and is designed to break at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator; breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handling.Without the actual sample a root cause can't be determined.The batch record for pn 312400 ln 8057519 was reviewed and there were no non-conformances related to broken ampoule during the manufacturing of this lot.No further action will be taken based on no adverse trend has been observed.Will continue to track & trend for this defect.H3 other text : see narrative below.
|
|
Event Description
|
Empty material (no solution).Customer sent to us the information below.What is the batch of product? the number 807519 is invalid (not found in our database) sorry for the misunderstanding, following batch: 8057519.
|
|
Search Alerts/Recalls
|