CAREFUSION 213, LLC 0113 CHLORAPREP ONE STEP; 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL Back to Search Results
Catalog Number 312400
Device Problems Defective Device (2588); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2020
Event Type  malfunction  
Event Description
Empty material (no solution).
 
Manufacturer Narrative
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.
 
Manufacturer Narrative
Based on no photograph or sample was received for analysis from the customer, the failure mode could not be confirmed.The ampoule is made of glass and is designed to break at a relatively low break force.When pressure is applied to the wings by a pinching force with the fingers, this activates the applicator; breaking the glass ampoule and releasing the chloraprep solution onto the foam tip.The most probable root cause can be attributed to post manufacturing handling, which can provide enough force/impact to activate and break the glass ampoule.Due to the nature of glass it is possible to have an activated applicator and/or broken ampoule if the applicator undergoes excessive handling.Without the actual sample a root cause can't be determined.The batch record for pn 312400 ln 8057519 was reviewed and there were no non-conformances related to broken ampoule during the manufacturing of this lot.No further action will be taken based on no adverse trend has been observed.Will continue to track & trend for this defect.H3 other text : see narrative below.
 
Event Description
Empty material (no solution).Customer sent to us the information below.What is the batch of product? the number 807519 is invalid (not found in our database) sorry for the misunderstanding, following batch: 8057519.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE STEP
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION 213, LLC 0113
1550 northwestern dr
el paso TX 79912
Manufacturer (Section G)
CAREFUSION, INC
75 n. fairview drive
vernon hills IL 60061
Manufacturer Contact
anna wehrheim
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key12914403
MDR Text Key286136469
Report Number3004932373-2021-00512
Device Sequence Number1
Product Code KXG
UDI-Device Identifier30382903124009
UDI-Public(01)30382903124009
Combination Product (y/n)Y
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Catalogue Number312400
Device Lot Number8057519
Date Manufacturer Received02/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-