Model Number CNS-6801A
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2021
Event Type  malfunction  
Event Description
The customer reported that their central nurse's station (cns) suddenly rebooted on two occasions.No harm or injury occurred.
 
Manufacturer Narrative
Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Manufacturer Narrative
Details of complaint: the customer reported that the central nurse's station (cns) suddenly rebooted on two occasions.No patient harm was reported.Investigation summary: the cns logs, dump files, and logsave files were sent in for analysis.In our analysis of these files, no abnormality was found in the os or hardware.Also, when the restart occurred, the screen application terminated abnormally, but the dump file was not created, and the cause could not be identified.We requested network capture of four cnss as well as captured data of the eg server, but they could not be obtained and resulted in investigation termination.The clear cause could not be determined because the requested information could not be obtained.Actions for the customer if the same event is reported again, please get the network capture logs directly related to the cns and eg server.The analysis concluded that no malfunction of the cns was observed.In order to understand the issue further, more data needed to be obtained.Service history for this serial number shows no related events.
 
Event Description
The customer reported that the central nurse's station (cns) suddenly rebooted on two occasions.No harm or injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-6801A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key12920221
MDR Text Key286165834
Report Number8030229-2021-02022
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921131640
UDI-Public04931921131640
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-6801A
Device Catalogue NumberPU-681RA
Was Device Available for Evaluation? No
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-