Model Number 87047
Device Problems Computer Operating System Problem (2898); Noise, Audible (3273)
Patient Problem Unspecified Heart Problem (4454)
Event Date 11/12/2021
Event Type  malfunction  
Manufacturer Narrative
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
 
Event Description
It was reported that after ablating in the left atrium with an intellanav mifi open-irrigated catheter, the equipment was brought back to right atrium for cavo-tricuspid isthmus (cti) line ablation.During cti line, the system lagged significantly almost to the point of freeze.Data was no longer transmitted while delivering radiofrequency (rf).The physician didn't observe any rf data due to system lag.During rf, with no direct sense data available, the physician observed a steam pop.The physician attributed the steam pop to the lag of the system.The procedure was successfully completed with no patient complications.The patient was stable.The flow rate was 30 ml.Direct sense impedance was temporarily unavailable due to system lag.No char was observed on the tip.The device was disposed of and therefore will not be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key12920567
MDR Text Key281665786
Report Number2134265-2021-15272
Device Sequence Number1
Product Code OAD
UDI-Device Identifier08714729938361
UDI-Public08714729938361
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150005/S017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2023
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0027489321
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-