BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number 87047 |
Device Problems
Computer Operating System Problem (2898); Noise, Audible (3273)
|
Patient Problem
Unspecified Heart Problem (4454)
|
Event Date 11/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported that after ablating in the left atrium with an intellanav mifi open-irrigated catheter, the equipment was brought back to right atrium for cavo-tricuspid isthmus (cti) line ablation.During cti line, the system lagged significantly almost to the point of freeze.Data was no longer transmitted while delivering radiofrequency (rf).The physician didn't observe any rf data due to system lag.During rf, with no direct sense data available, the physician observed a steam pop.The physician attributed the steam pop to the lag of the system.The procedure was successfully completed with no patient complications.The patient was stable.The flow rate was 30 ml.Direct sense impedance was temporarily unavailable due to system lag.No char was observed on the tip.The device was disposed of and therefore will not be returned.
|
|