The suspect device has not been returned to olympus for evaluation.The legal manufacturer completed the investigation.The dhr for this device were not reviewed because no lot number was provided.Olympus ships products that are manufactured according to all applicable procedures and meet final product release criteria.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is undetermined.The following information is provided in the instructions for use, rev an: "when the ultrasound is activated, do not contact tissue with the tip or sides of the probe; tissue damage could result." (additional tip#9, page 23); "do not contact mucosal tissue with the tip of the activated probe as tissue damage may occur." (warning, page 23); and "the tip of the probe must be kept visible at all times when the transducer is activated." (caution, page ).
|
The customer reported to an olympus representative, during a cystolitholapaxy with the shockpulse probe, the patient's bladder was perforated.The probe did not come in direct contact with the mucosal tissue.The probe did come in direct contact with the stone and pressure was applied to the stone with the probe.The size of the stone was 3-4 cm.There were not any procedural or anatomical challenges that could have caused or contributed to the bladder perforation.The power setting during the procedure was high.The vacuum pressure was 50 mmhg.Continuous irrigation was used with normal saline solution.No devices malfunctioned during the procedure.The probe was discarded after use and not available for return.The perforated bladder was treated with catheter placement post-procedure and will remain for two weeks.The patient is reported as 'fine'.
|