• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-Q260
Device Problem Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/19/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to any of the olympus locations.Therefore, olympus could not investigate the subject device.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus (b)(4) and similar past cases, since it is undeniable that the subject device was used with a foreign material on the air/water nozzle, omsc surmised that this phenomenon was attributed to the following.There was a difference between the reprocessing method performed by the user facility and the reprocessing method recommended by the instruction manual.The staff of the user facility had not been trained sufficiently on the handling and reprocessing the subject device according to the instruction manual.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the unspecified therapeutic gastroscopy using the subject device, it was found the air supply of the subject device was insufficient, which was caused by clogging of the air/water nozzle due to protein coagulation after bleeding of patient.The subject device had been repaired on site by the engineer of the user facility.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12931036
MDR Text Key286377979
Report Number8010047-2021-15526
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGIF-Q260
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-