We checked the returned and confirmed that the ccd module defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the air/water nozzle loosened, the insertion flexible tube (ift) leaky, the segment loosened, the u/d and the r/l pulley wire worn out, and the remote control buttons cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|