ENCORE MEDICAL L.P RSP; RSP BASEPLATE, 30MM, W/P2 COATING Back to Search Results
Model Number 508-32-204
Device Problems Break (1069); Improper or Incorrect Procedure or Method (2017); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 11/08/2021
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was reported as broken center screw.The previous surgery and the surgery detailed in this event occurred 1 year and 3 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements and showed no non-conforming material reports (ncmr) associated with the main contributor component listed in the event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to broken center screw there are no indications of a product or process issue affecting implant safety or effectiveness.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - broken center screw.Surgeon states the patient increased activity level prior consent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
RSP BASEPLATE, 30MM, W/P2 COATING
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
kiersten soderman
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key12936982
MDR Text Key281828892
Report Number1644408-2021-01328
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912144391
UDI-Public(01)00888912144391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112069
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/06/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number508-32-204
Device Catalogue Number508-32-204
Device Lot Number769P1843
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
506-03-126 LOT 834C1698; 506-03-126 LOT 834C1712; 508-36-101 LOT 869C2954; 509-00-036 LOT 378P1265
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexMale
-
-