Sample from the patient was tested as part of the investigation.The ft4 iii result from a cobas e801 was 8 mg/dl with a data flag and using a cobas e411 was 10 ng/dl.The ft4 ii result using a cobas e601 was 8 ng/dl with a data flag and using a cobas e411 was 28 ng/dl the difference in the results could be explained by the different calibrations between the cobas e411 and the cobas e601.No significant change in signal values was observed in the samples with reagent with streptavidin blocking agent.With the three different ft4 assay generations (ii, iii and iv), similar very high ft4 values were generated.Further clarification of the discrepancy is not possible with the available methods and the current state of the art.The investigation did not identify a product problem.The cause of the event could not be determined.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination, and other findings.
|