Additional information: the device has not been returned for analysis due to this being a literature review, and after multiple attempts to gather more information from the reporter, no additional information has been received.The investigator determined a device history record (dhr) review is not possible for this complaint, as attempts at gathering the device serial/lot number were unsuccessful.A review of the device labeling notes the following: the current overstitch¿ endoscopic suturing system (ess) instructions for use (ifu) addressed the known and potential event of "suture assembly-suture broke during procedure; vessel damage / bleeding and multiple symptoms" ess - " as follows: warning: do not introduce the device with the needle body in its open position.Warning: ensure appropriate suture slack has been created for desired suture path and pattern.Advance anchor exchange and/or manipulate endoscope to create suture slack.Anchor exchange will not install anchor onto the needle body: i.Ensure there is sufficient suture slack and the suture outside endoscope is not entangled.Ii.Ensure anchor exchange is properly positioned in alignment tube of needle driver.Warnings: only physicians possessing sufficient skill and experience in similar or the same techniques should perform endoscopic procedures.Note: ensure suture is not tangled after removal from the cartridge.Caution: do not use when valve covers are closed as suture drag will be increased.Caution: if resistance is encountered when advancing the anchor exchange through the working channel of the endoscope, reduce the endoscope angulation until the device passes smoothly.Ensure that there is sufficient space for the needle to open.The labeling is adequate as it addresses the reported complaint.The occurrence of this reported complaint and product will be monitored as appropriate.
|