The mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: unit received with the lock having both rotational and lateral movement and a residue buildup is present.Upon disassembly repair noted the index knob and the lock will need new components added to replace worn internal parts; unit was machined to have heli-coils added to large starburst threads.The unit needs repair, pm and replacement of worn, damaged and missing parts.General maintenance and cleaning.Root cause analysis: the reported complaint was confirmed from the evaluation.Parts of the device are worn and damaged.Evaluation showed that the lock had movement and a residue buildup is present.The definite root cause cannot be reliably determined.The most probable root cause is wear and tear / improper handling.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|