Model Number A2000
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the mayfield 2000 skull clamp (a2000) is not locking properly.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
 
Manufacturer Narrative
Mayfield 2000 skull clamp (a2000) was returned for evaluation: failure analysis & root cause: complaint confirmed via inspection of the unit.Unit received with the hex bushing worn and the lock could still move after the unit was locked down and needs to be replaced.General maintenance and cleaning required at this time.All worn components replaced with new parts.Additionally, unit required preventive maintenance and replacement of worn parts as a result of normal wear and tear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD 2000 SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key12975426
MDR Text Key284816189
Report Number3004608878-2021-00683
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253655
UDI-Public10381780253655
Combination Product (y/n)N
PMA/PMN Number
K932807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA2000
Device Catalogue NumberA2000
Device Lot Number017(2001)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2021
Date Manufacturer Received11/19/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-