MAUDE Adverse Event Report: AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. AUTOSTAINER LINK 48; AUTOMATED SLIDE STAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD. AUTOSTAINER LINK 48; AUTOMATED SLIDE STAINER Back to Search Results
Model Number AS480
Device Problems Fluid/Blood Leak (1250); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/31/2021
Event Type  malfunction  
Manufacturer Narrative
The dispense check valve malfunction may potentially impact staining performance.No erroneous staining result was reported by the customer in connection with this incident.No patient or user harm was indicated.Patient information has not been provided by the user.
 
Event Description
The customer reported "reagent needle dripping." the field service engineer replaced the dispense check valve causing this malfunction.The instrument is fully operational and performs as intended.There was no delay in generating staining results.No direct or indirect patient harm or user harm have been reported.
 
Event Description
The customer reported "reagent needle dripping." the field service engineer replaced the dispense check valve causing this malfunction.The instrument is fully operational and performs as intended.There was no delay in generating staining results.No direct or indirect patient harm or user harm have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOSTAINER LINK 48
Type of Device
AUTOMATED SLIDE STAINER
Manufacturer (Section D)
AGILENT TECHNOLOGIES SINGAPORE (INTL.) PTE LTD.
no.1 yishun avenue 7
singapore north east, 76892 3
SN  768923
Manufacturer (Section G)
SHANDON DIAGNOSTICS LIMITED
tudor road
manor park
runcorn, WA7 1 TA
UK   WA7 1TA
Manufacturer Contact
bryce portier
1834 state highway 71 west
cedar creek, TX 78612
4082185467
MDR Report Key12980838
MDR Text Key283316330
Report Number3003423869-2021-00306
Device Sequence Number1
Product Code KPA
UDI-Device Identifier05700572035497
UDI-Public05700572035497
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 12/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAS480
Device Catalogue NumberAS48030
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/31/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-