Model Number AS480 |
Device Problems
Fluid/Blood Leak (1250); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/31/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The dispense check valve malfunction may potentially impact staining performance.No erroneous staining result was reported by the customer in connection with this incident.No patient or user harm was indicated.Patient information has not been provided by the user.
|
|
Event Description
|
The customer reported "reagent needle dripping." the field service engineer replaced the dispense check valve causing this malfunction.The instrument is fully operational and performs as intended.There was no delay in generating staining results.No direct or indirect patient harm or user harm have been reported.
|
|
Event Description
|
The customer reported "reagent needle dripping." the field service engineer replaced the dispense check valve causing this malfunction.The instrument is fully operational and performs as intended.There was no delay in generating staining results.No direct or indirect patient harm or user harm have been reported.
|
|