• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN PROTEGE RX CAROTID STENT SYSTEM; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN PROTEGE RX CAROTID STENT SYSTEM; STENT, CAROTID Back to Search Results
Catalog Number SEPX-8-6-40-135
Device Problems Entrapment of Device (1212); Difficult to Remove (1528)
Patient Problem Insufficient Information (4580)
Event Date 12/05/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to use the spider fx embolic protection device along with protege rx carotid stent during procedure to treat severely calcified lesion in the proximal right common carotid artery.The vessel exhibited 90% stenosis.The vessel was pre-dilated.Ifu was followed.Angiography was performed.There was an issue identified of preparing the spider and stent of appropriate specifications.It was reported that the spider was hung on the stent when it was being taken out of the body and could not be taken out.The stent did pass through a previously deployed stent.No resistance was encountered advancing device.The procedure was completed by taking out the spider and stent by carotid artery peeling.The devices were surgically removed from the patient.No vessel damage noted.No patient injury.
 
Manufacturer Narrative
Product analysis: the spider fx device was returned to medtronic investigation lab for evaluation.The deployed stent was returned along with a basket from a spider fx device inside a small ziplock bag in a biohazard bag in its shelf carton.The deployed protégé rx stent was received along with a basket from a spider fx device.The stent was confirmed as 40mm.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGE RX CAROTID STENT SYSTEM
Type of Device
STENT, CAROTID
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key12984462
MDR Text Key286766220
Report Number2183870-2021-00454
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P060001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2023
Device Catalogue NumberSEPX-8-6-40-135
Device Lot NumberB262575
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-