Quality controls were run on both meters and were acceptable.The meter and strips have been requested for return.All testing with the returned strips produced acceptable results and no strip defects were observed.Testing summary: 12 strips returned from strip lot# 480034 (vial 00754).The strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.No obvious reagent discoloration.12 strips returned from lot# 480034 (vial 00392).The strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.No obvious reagent discoloration.Three strips returned from strip lot# 480034 (vial 01358).The strip vial, desiccant, and vial cap were inspected and were acceptable in appearance.No obvious reagent discoloration.A routine retention testing process has been implemented.Valid retention results are available for all lots in the event a strip lot is alleged in a complaint.All retention data is reviewed on a monthly basis once testing is complete.The investigation is ongoing.
|
There was an allegation of questionable glucose results from two accu-chek inform ii meters.At 17:58 using meter serial number (b)(4), the result was 500 mg/dl.At 17:59 using meter serial number (b)(4), the result was 105 mg/dl.At 18:07 using meter serial number (b)(4), the result was 243 mg/dl.
|