The biosense webster, inc.Product analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.Manufacturer's reference number: (b)(4).
|
The device evaluation was completed on 16-dec-2021.It was reported that a patient underwent a paroxysmal atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter and an unwanted pacing issue occurred.It was reported that a pacing spike was displayed on the carto 3 system when not pacing the catheter.It just looked like a pacing spike but nothing was being stimulated."no stim channels were connected.It was a bloom".The cable was replaced without resolution.The catheter was replaced, and the issue resolved.The case continued without further incident or harm.The carto 3 system is operating per specs and was not responsible for the product issue.No adverse patient consequences were reported.Device evaluation details: the device was returned to biosense webster inc.(bwi) for evaluation.A visual inspection and electrical test of the returned device were performed in accordance with bwi procedures.Visual analysis of the returned device revealed that no damage or anomalies were observed on the device.An electrical test was performed, and no electrical issues were found.A manufacturing record evaluation was performed for the finished device 30648661l number, and no internal action related to the complaint was found during the review.The electrical issue reported by the customer could not be replicated during the product investigation; other issues or circumstances may have occurred during the usage of the device that compromised its performance.The instructions for use contain the following recommendation: ecg noise is typically generated as the result of the improper connection of the body surface ecg patch to the patient.This noise is the most significant during ablation.To resolve this situation, verify the proper connection.It is recommended to turn off the notch filter for this verification.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|