MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
Model Number 5540030 |
Device Problems
Break (1069); Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from healthcare professional via field representative regarding a patient with symptoms of lumbar disc hern iation involved in the lumbar interbody decompression and fusion fixation procedure.It was reported that intra-operatively, product was not tight enough, a little loose after self-detaching, and the screw plug was replaced.Screw broke and no fragments left in the patient's body.There were no patient symptoms reported.There were no further complications reported as a result of this event.Patient is alive and no injury.On 2021-nov-24, received additional information that there was delay in overall procedure time for 1 minute.
|
|
Manufacturer Narrative
|
H3: device evaluated summary- visual inspection confirmed the only the upper break off portion of the set screw was returned.Unable to determine root cause of issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|