Catalog Number 110003452 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Concomitant medical products: it is unknown which cup was removed from the patient: cat#110010243 g7 osseoti 3 hole shell 50mm d lot#65095363, cat#110010243 g7 osseoti 3 hole shell 50mm d lot#65028595.Customer has indicated that the product will not be returned to zimmer biomet for investigation as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip procedure the rod for inserter sheared off at threads leaving portion of threads stuck in acetabular implant which caused damage to implant.Forced to replace implant.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Visual examination of the provided pictures identified the inserter to be fractured with a portion of the inserter still assembled to the shell.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|