DORNOCH DUO FLUID CART - REFURBISHED; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Sparking (2595)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/30/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the cart would spark when plugged into the wall after surgery.There was no patient involvement.No additional consequences have been reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up is being submitted to relay additional information.The results of the investigation are as follows: the event is confirmed.Review of the most recent repair record determined that the cart would spark when plugged into the wall receptacle, and the power cord plug prongs were damaged.The power cord was replaced and resolved the reported issue.Review of the previous repair record identified unrelated repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|