Catalog Number A103200000000 |
Device Problems
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
N this event it was reported that an x-smart plus contra angle won't hold files; no injury resulted.
|
|
Manufacturer Narrative
|
Multiple unsuccessful attempts were made to obtain the device for evaluation.
|
|
Manufacturer Narrative
|
Multiple unsuccessful attempts were made to obtain the device for evaluation.
|
|
Manufacturer Narrative
|
Device for this complaint was received and after inspection/investigation found received = 1x x-smart plus bearing retainer h1033c1051113.1x x-smart plus cartridge h1033c1051015.1x x-smart plus drive shaft h1033c1051022.1x x-smart plus head cap h1033c1051050.1x x-smart plus wave washer h1033c427a360.X-smart plus cartridge.Various mechanical problem.Inside the cartridge moves (when a file is fixed).
|
|