OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/19/2021
Event Type  malfunction  
Event Description
This device was returned for repair by the customer for a leak observed during reprocessing.There is no patient involvement and no harm reported to any patient.During the time of the device evaluation, it was observed that the forceps cover glue was peeling.This medwatch is being submitted for the reportable malfunction of forceps cover glue peeling observed during evaluation.
 
Manufacturer Narrative
The device is returned and an evaluation completed for it.Upon inspection and testing, it was observed that the forceps cover glue was peeling.Additionally, the k-lever is leaking when manipulated and the distal end rubber cover glue is chipped.The scope connector is loose.The user¿s complaint of leak was confirmed.Evaluation is ongoing.Supplemental report(s) will be submitted when any relevant new information is available.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, external damage was observed at the distal end.However, the definitive root cause of the reported issue could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿do not apply shock to the distal end of the insertion section, particularly the ultrasound transducer and the objective lens surface at the distal end.Visual abnormalities may result.¿ it was confirmed that the subject device met its standards at the time it was shipped.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13011503
MDR Text Key282290414
Report Number8010047-2021-16182
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356339
UDI-Public04953170356339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/15/2021
Supplement Dates Manufacturer Received01/05/2022
Supplement Dates FDA Received01/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-