• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE RELINE SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 13550000
Device Problems Expulsion (2933); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
No product was received although a radiograph was provided confirming the alleged lock screw expulsions.The patient was reported to have followed post op activity instructions and had not experienced any falls or other accidents.It is unknown whether the backed out screws from a previous revision were reused or replaced.While a root cause for the issue could not be confirmed review of the information provided would identified a previous revision occurred for lock screw back outs and suggests possible thread damage incurred, rod interference or excessive loading as possible cause or contributors.No additional investigation can be completed.Label review: "potential adverse events and complications - as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries." potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation." "all lock screws should be final-tightened with the counter-torque and torque t-handle.Do not final-tighten through compression instruments." "cross connectors are designed specific to the rod diameter and cannot be used on the tapered section of tapered rods.If using cross connectors on tapered rods, only attach them on constant diameter rod sections.Care should be taken to insure that all components are ideally fixated prior to closure." "single use/do not re-use: reuse of a single use device that has come in contact with blood, bone, tissue or other body fluids may lead to patient or user injury.Possible risks associated with reuse of a single use device include, but are not limited to, mechanical failure, material degradation, potential leachables, and transmission of infectious agents." device not returned.
 
Event Description
On a unknown date the patient was reported to be hearing the sound of metal coming from his back.It was then discovered during a follow with his physician that two connector lock screws had completely expelled from the connector and were floating loosely in the patient.On a unknown date a revision occurred to address the loosened screws.During the revision the expelled lock screw were replaced.The patient is reported to be doing well post revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key13011877
MDR Text Key287277386
Report Number2031966-2021-00186
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517551702
UDI-Public887517551702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number13550000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
-
-