• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 08P1934
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed falsely elevated alinity c magnesium results generated on an alinity c processing module for one patient.Sid (b)(6) initial result = 2.713 mmol/l, repeat results = 0.618 mmol/l and 0.618 mmol/l.There was no impact to patient management reported.
 
Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.Ticket and trending review did not identify any trends for the complaint issue or likely cause lot number.Device history record review did not identify any non-conformance's or deviations with the likely cause list number.File sample analysis was not performed because as part of troubleshooting, the sample was retested, and the results were lower.Quality control results were within expected ranges during the complaint issue, indicating the assay was performing as expected.Labeling was reviewed and found to adequately address the issue under review.Based on the investigation, no systemic issue or deficiency with the alinity c magnesium reagent lot 27374ud00 was identified corrected information in g1: contact office first name, contact office last name, contact office address 1, contact office address 2, contact office city, contact office country, contact office postal code, contact office phone number, contact office fax number, contact office e-mail.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Type of Device
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford, IL N39 E-932
2246682940
MDR Report Key13013961
MDR Text Key284142781
Report Number3005094123-2021-00244
Device Sequence Number1
Product Code JGJ
UDI-Device Identifier00380740176532
UDI-Public00380740176532
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K181748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2022
Device Model Number08P1934
Device Catalogue Number08P19-34
Device Lot Number27374UD00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ALNTY C PROCESSING MODU, 03R67-01, AC01130; ALNTY C PROCESSING MODU, 03R67-01, AC01130
-
-