• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG DEXTRUS 53 MODEL 4136 US 644136-103; PACING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG DEXTRUS 53 MODEL 4136 US 644136-103; PACING LEAD Back to Search Results
Model Number 358754
Device Problems High impedance (1291); Over-Sensing (1438); Pacing Intermittently (1443)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2021
Event Type  malfunction  
Event Description
This rv lead was explanted due to impedances of greater than 2000 ohms, oversensing and pacing inhibition.No adverse patient events were reported.
 
Manufacturer Narrative
The lead is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXTRUS 53 MODEL 4136 US 644136-103
Type of Device
PACING LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13017993
MDR Text Key282343271
Report Number1028232-2021-06751
Device Sequence Number1
Product Code NVN
UDI-Device Identifier04035479143213
UDI-Public04035479143213
Combination Product (y/n)N
Reporter Country CodeDE
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2009
Device Model Number358754
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/18/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-