Catalog Number 261221 |
Device Problem
Fail-Safe Did Not Operate (4046)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
A facility reported a perforator failed to disengage.It is unknown if there was patient involvement; however, no patient injury or surgical delay has been reported.
|
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
The perforator was returned for evaluation: dhr - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.The unit received for this complaint was unused, in its original sealed packaging.Ifu testing was performed with no observed anomalies.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release: the unit was found to perform as intended and fulfilled the acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
|
|