Catalog Number 295400-001 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/01/2021 |
Event Type
malfunction
|
Event Description
|
The freedom hospital ac power supply is a component that enables the freedom driver
to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom hospital ac power supply had a damaged connector.The freedom hospital ac power supply was not supporting a patient.
|
|
Manufacturer Narrative
|
The freedom hospital ac power supply will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4).
|
|
Manufacturer Narrative
|
During visual inspection, damage to the connector was confirmed.Additionally, one of the four power supply rubber feet were missing and physical damage was observed on the power brick.The investigation confirmed the customer-reported issue.The observed damage was mostly likely due to accidental rough handling.The power supply's functionality was not able to be evaluated due to the physical damage.This issue will be monitored and trended as part of the customer experience process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
|
|