Model Number 397002-001 |
Device Problems
Battery Problem (2885); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/03/2021 |
Event Type
malfunction
|
Event Description
|
The customer, a syncardia authorized distributor, reported the companion 2 driver exhibited an emergency battery error alarm.The companion 2 driver was not supporting a patient.
|
|
Manufacturer Narrative
|
The companion 2 driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.(b)(4) initial.
|
|
Manufacturer Narrative
|
The customer-reported issue was confirmed upon review of the patient file which revealed that the driver had experienced an emergency battery error alarm and replicated during incoming functional testing when the driver exhibited an emergency battery error alarm.Additionally, during testing, the emergency battery was unable to power the driver.The root cause of the customer-reported emergency battery alarm was determined to be a malfunction of the emergency battery.The battery was found to have a permanent fault status which prohibited the battery from charging or discharging.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.Ce 5720 comp(b)(4) follow-up report 1.
|
|