SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Mechanical Problem (1384); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver was defective, however there was no additional information provided.There was no reported adverse patient impact.The customer also reported that the patient was subsequently changed to a backup driver.
 
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Manufacturer Narrative
The driver's alarm history was reviewed and revealed a 2d alarm code which is recorded when the secondary motor is engaged.Visual inspection of the interior components of the driver revealed that the secondary motor was out of bottom dead center (bdc) position, the scotch yoke was cracked and the primary motor was binding when turned by hand and was covered in debris.Upon review of the driver's alarm history and visual inspection, it is reasonable to concluded that the freedom driver exhibited an alarm which caused the patient to switch drivers.The root cause of the fault alarm was accumulated debris which caused the primary motor to bind up, leading to the secondary motor operation and annunciated alarm as designed.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key13025094
MDR Text Key284891438
Report Number3003761017-2021-00236
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age57 YR
Patient SexFemale
-
-