• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100R
Device Problems Loose or Intermittent Connection (1371); Intermittent Communication Failure (4038)
Patient Problems Head Injury (1879); Failure of Implant (1924); Unspecified Tissue Injury (4559)
Event Date 03/30/2022
Event Type  Injury  
Event Description
The recipient reported is experiencing loss of lock following a head trauma incident.External equipment was exchanged, however the issue did not resolve.Revision surgery will be scheduled.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.A review of the device history record was completed and no anomalies were noted.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.The recipient¿s device will not return to advanced bionics for analysis.The recipient has recovered well.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key13027635
MDR Text Key282418272
Report Number3006556115-2021-01969
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/01/1998
Device Model NumberAB-5100R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/04/1996
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age50 YR
Patient SexFemale
-
-