Model Number EPK-I7010
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2021
Event Type  malfunction  
Event Description
On (b)(6), the distributor purchased the epk-i7010 s/n: (b)(4) from beijing stone jiahe company.On (b)(6) 2021, epk-i7010 s/n: (b)(4) arrived at office.On (b)(6) 2021, the distributor opened the box and inspected the processor found that horizontal lines on the image of touch panel.They tried 5 times; the defect happened at 1st (2 mins), 4th time, 5th time (3 mins after power-on for 5 mins).This event occurred at the time of before use.There was no report of patient harm.
 
Manufacturer Narrative
If additional information becomes available, a supplemental report will be filed with the new information.This device is not distributed in us so that unique identifier is blank.This device is classified as import for export, therefore 510k is not applicable.Model epk-i7010-us is available in the usa with a 510k number k182004.
 
Manufacturer Narrative
Evaluation summary: the defect happened at distribuor office.And before use.The processor has been replaced.Correction information: h6: coding changed, based on the investigation result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO PROCESSOR
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key13040456
MDR Text Key287035173
Report Number9610877-2021-01924
Device Sequence Number1
Product Code PEA
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEPK-I7010
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-