BOSTON SCIENTIFIC CORPORATION EMBLEM MRI S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number A219
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pocket Stimulation (1463); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Device Overstimulation of Tissue (1991); Electric Shock (2554)
Event Date 10/01/2021
Event Type  Injury  
Event Description
It was reported that a patient with this subcutaneous implantable cardioverter defibrillator (s-icd) was hospitalized after receiving multiple inappropriate shocks.Review of the episode noted oversensing of noise contributing to the delivery of inappropriate therapy.The patient may have been experiencing muscle stimulation from their s-icd as well when they were at the hospital.The sensing vector of the s-icd was reprogrammed and the device remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM MRI S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13048851
MDR Text Key282549688
Report Number2124215-2021-31565
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00802526584404
UDI-Public00802526584404
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P110042/S058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/10/2021
Device Model NumberA219
Device Catalogue NumberA219
Device Lot Number245544
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/19/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-