• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM; SPINAL VERTEBRAL BODY REPLACEMENT DEVICE Back to Search Results
Model Number 436013C
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Event Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having cervical corpectomy spinal therapy at c4.It was reported that tooth on worm gear inside of centerpiece broke off during expansion.Prevented surgeon from using the cage to distract under load.The part came in contact with the patient.There was no patient symptom reported.There were no further complications reported regarding the event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: part# 436013c lot# ca19h095 visual and optical inspection confirmed a couple of the teeth of the centerpiece implant have been damaged/broken.Functional inspection with a sample inserter confirmed the implant would still expand and close but would not expand when excessive force was placed on it.This type of damage is consistent with excessive force placed on the implant causing it to strip out.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T2 STRATOSPHERE¿ EXPANDABLE CORPECTOMY SYSTEM
Type of Device
SPINAL VERTEBRAL BODY REPLACEMENT DEVICE
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key13050134
MDR Text Key288800030
Report Number1030489-2021-01576
Device Sequence Number1
Product Code MQP
UDI-Device Identifier00643169663138
UDI-Public00643169663138
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number436013C
Device Catalogue Number436013C
Device Lot NumberCA19H095
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
Patient SexMale
Patient Weight95 KG
-
-