Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.Mayfield composite series skull clamp (a3059) was returned for evaluation.Device history record (dhr): the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis: complaint confirmed via inspection of the unit.Unit received with the mayfield 2 lock having up and down movement and the teeth were grinding when rotated.General maintenance and cleaning has been performed.Root cause analysis: the reported complaint confirmed via inspection of the unit.The mayfield 2 lock had movement and the teeth were grinding when rotated.Unit required preventive maintenance as a result of routine wear and tear.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|