• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3852
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported that the balloon ruptured.A 15mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the balloon ruptured.The procedure was completed with another of the same device.There was no patient injury.
 
Event Description
It was reported that the balloon ruptured.A 15mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the balloon ruptured.The procedure was completed with another of the same device.There was no patient injury.
 
Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A visual examination of the balloon found no tears or holes in the balloon material.An attempt was made to inflate the balloon which failed.The failure to inflate was consistent with the large build-up of solidified media present inside the balloon.The device was soaked in a water bath at 37 degrees celsius in order to soften the solidified media.The returned device was attached to an encore inflation unit and positive pressure was applied in an attempt to inflate the balloon.Liquid was observed to be leaking from a balloon pinhole located at the mid-body of the balloon.A microscopic examination of the balloon material and markerband identified no issues which could potentially have contributed to the pinhole leak.No damage was observed to any of the blades.All blades are fully bonded onto the balloon.A visual and tactile examination found no issues with the hypotube of this device.The markerbands and tip and of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13076243
MDR Text Key282713369
Report Number2134265-2021-16165
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/06/2023
Device Model Number3852
Device Catalogue Number3852
Device Lot Number0027793859
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-