• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2021
Event Type  malfunction  
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs.While troubleshooting, fse found the main arm z-axis movement was erratic and the sample probe sticking up due to lack of lubrication which would not allow consistent tip detachment.Fse cleaned and lubricated the main arm z-axis lead screw, guide, and nozzle bushing which resulted in a smoother consistent movement.Fse successfully performed a function test for the tip detachment without errors.Fse performed quality control run without error and within acceptable range.No further action required by field service.The aia-2000 instrument is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-2000 operator's manual under appendix 4: error messages states the following: [2061] tip detachment failure by main arm cause : a tip was detected by tip detachment check.If retry fails, the measurement result will be flagged (mf flag).Solution : contact tosoh service center or local representatives.The most probable cause of the reported event was due to lack of lubrication causing a sticky main arm sample nozzle.
 
Event Description
A customer reported getting error message "2061 tip detachment failure by main arm" on the aia-2000 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
bernadette o connell
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368143
MDR Report Key13082431
MDR Text Key286672799
Report Number3004529019-2021-00088
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-