Model Number S-60-040-120-P6
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation has not yet been completed.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during preparation of the supera self-expanding stent (ses), the nosecone came off the device.There was no forced manipulation, it just easily came off.Therefore, the device was exchanged for a new supera.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was returned for analysis.The reported material separation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that inadvertent mishandling during unpackaging/stylet removal and/or during preparation for use resulted in the reported tip separation/noted inner member separation.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13098136
MDR Text Key286758908
Report Number2024168-2021-12228
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08717648211799
UDI-Public08717648211799
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberS-60-040-120-P6
Device Catalogue NumberS-60-040-120-P6
Device Lot Number1022462
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-